Exo User Manual Chlorophyll Calibration Rhodamine
- Exo User Manual Chlorophyll Calibration Rhodamine Guide
- Exo User Manual Chlorophyll Calibration Rhodamine 1
- Exo User Manual Chlorophyll Calibration Rhodamine Free
Calibration menus quickly guide you through the calibration process and provide SmartQC verification of successful calibration and worksheets. EXO Sondes: What's in the Box When it Arrives? In this unboxing video, Tim Finegan, YSI, shows you exactly what ships with every order of an EXO water quality monitoring sonde. Experienced user with the steps. Required to make your YSI 6-Series. Product operational. For full instructions, please refer to. The 6-Series User Manual. Optical probes (6025 chlorophyll, 6130. Rhodamine, 6131 phycocyanin blue-green. Algae, 6132 phycoerythrin blue-green algae, 6136 turbidity, and 6150 ROX DO): Insert. EXO chlorophyll readings show excellent linearity on serial dilution of a surrogate solution of Rhodamine WT (R 2 0.9999) and this should ensure relative accuracy of field chlorophyll readings, i.e., a chlorophyll reading of 100 units will represent twice the algal content of water with a chlorophyll reading of 50 units. Self-monitoring system reports the status of the instrument, flags the data, and shows the user where the problem is with assistance on how to solve the issue; User-scheduled calibration and maintenance intervals indicate when they are due; Guided and semi-automated calibration routines lead the user through the calibration process. The EXO handheld is a portable weatherproof interface for the EXO sondes that also comes with a number of unique features. These features include GPS, temperature-compensated barometer, wet-mate wireless connector, Bluetooth communications, and 2 GB of storage. Besides the handheld, the EXO sonde has multiple data outputs.
*Specifications indicate typical performance and are subject to change. All sensors have a depth rating to 250 m (820 ft), except shallow and medium depth sensors, ammonium, chloride, and nitrate. Accuracy specification is attained immediately following calibration under controlled and stable environmental conditions. Performance in the natural environment may vary from quoted specification.
w.i.g. = whichever is greater
EXO sensors are not compatible with YSI 6-Series sondes, sensors, or handheld.
1Outputs of specific conductance (conductivity corrected to 25˚C) and total dissolved solids are also provided. See Calculated Parameters and footnote 11.
2 Temperature accuracy traceable to NIST standards
3Accuracy specifications apply to conductivity levels of 0 to 100,000 μS/cm.
4 Relative to calibration gases
5 When transferred from air-saturated water to stirred deaerated water
6 When transferred from water-saturated air to Zobell solution
7Within the environmental pH range of pH 4 to pH 10.
8On transfer from water-saturated air to rapidly stirred air-saturated water at a specific conductance of 800 μS/cm at 20˚C; T63<5 seconds on transfer from water-saturated air to slowly-stirred air-saturated water.
9 Performance based on 3-point calibration done with YSI AMCO-AEPA standards of 0, 124, and 1010 FNU. The same type of standard must be used for all calibration points.
10 Values are automatically calculated from conductivity according to algorithms found in Standard Methods for the Examination of Water and Wastewater (Ed. 1989).
Unpacking and Inspection
Inspect the outside of the shipping
container(s) for damage. If you see any
damage, contact your shipping carrier
immediately.
Remove the equipment from the shipping
box. Some parts are loose in the box, so check
the packing material carefully. Check off all
items on the packing list and inspect them for
damage.
Items Needed
•
Sonde
• Probes
• Calibration cup
• Probe guard
• Batteries (if sonde has battery compartment)
• DI water
• Flathead screwdriver
• Calibration standards
• Communication or field cable
If using 6562 DO probe:
• 6570 maintenance kit
• 5775 membrane kit
Exo User Manual Chlorophyll Calibration Rhodamine Guide
• Razor blade
6-Series Sonde
Quick Start Guide
Preparing Sonde for Use
To prepare the sonde for operation, you
will need to:
• Install a new membrane on the YSI 6562
DO Probe, if you are using this item.
• Install probes (sensors) into the connectors
on the sonde bulkhead.
• Provide power for the sonde through
batteries or line power.
• Connect a field cable or PC communication
cable.
It is important to calibrate all
sensors prior to using the sonde for field
measurements. Refer to Section 2.6.2 of the
6-Series User Manual.
Installing Membrane on
6562 DO Probe
(Skip to next section if you are not using
this probe)
The 6562 Rapid Pulse
™
Dissolved Oxygen
probe is shipped with a protective dry
membrane on the sensor tip. Remove the
O-ring and membrane. Handle the probe
with care. It is important not to scratch or
contaminate the sensor tip.
The DO sensor on the 600R/QS cannot
be removed.
Open the 6562 DO Probe Kit and follow these
steps:
1. Prepare the electrolyte solution as specified
on the bottle.
2. Put/keep the protective cap on the
connector end of the probe to protect it from
contamination.
3. Hold the probe in a vertical position and
apply a few drops of KCl solution to tip. The
fluid should completely fill the small moat
around the electrodes—and create a large
meniscus on the probe face. Be sure no air
bubbles are stuck to the face of the sensor. If
necessary, shake off electrolyte and start over.
4. Secure a membrane between your left
Exo User Manual Chlorophyll Calibration Rhodamine 1
thumb and the probe body. Touch membrane
only at the ends.
5. With the thumb and forefinger of
your right hand, grasp the free end of the
membrane and gently stretch it up, over,
and down the other side of the sensor. The
membrane should be stretched below the
ridge where the O-ring seats so a good seal is
achieved.
6. Secure the end of the membrane under the
forefinger of your left hand.
7. Roll the O-ring over the end of the probe,
being careful not to touch the membrane
surface with your fingers.
8. If bubbles or wrinkles are present, remove
the membrane and repeat steps 3-7.
9. Trim excess membrane with a razor blade.
Rinse off any excess KCl solution, but be
careful not to get any water in the connector.
Refer to Section 2.3 of the 6-Series User
Manual for complete instructions and photos.
Installing Probes
Before installing any probe into the
sonde bulkhead, be sure that the probe port
is free of moisture. Use a Kimwipe
™
or lint-
free cloth to dry.
1. Remove the calibration cup from your
sonde, if necessary.
2. Using the long end of the probe
installation tool supplied in the 6570
Maintenance Kit, remove the port plug from
the port.
3. Some probes can only be installed into
specific ports. Refer to Section 2.3 of the 6-
Series User Manual for port designations.
4. Apply a thin coat of lubricant (Krytox
™
)
to the O-rings on the connector side of the
probe.
Make sure there are NO contaminants
between the O-ring and the probe.
Contaminants will cause leaks when the
sonde is deployed.
When installing probes, be
careful not to cross-thread the
probe nuts. If you meet resistance
while finger-tightening, STOP,
back out, and begin again.
This document will assist an
experienced user with the steps
required to make your YSI 6-Series
product operational.
For full instructions, please refer to
the 6-Series User Manual.
5. Optical probes (6025 chlorophyll, 6130
Rhodamine, 6131 phycocyanin blue-green
algae, 6132 phycoerythrin blue-green algae,
6136 turbidity, and 6150 ROX DO): Insert
the probe in the appropriate optical port,
seating the pins of the two connectors. Hand-
tighten the probe nut to the bulkhead. Finish
tightening the probe using the short end of the
installation tool. Do not over-tighten.
6. 6562 Rapid Pulse DO, 6560 conductivity/ Sharp vl ah151u manual.
temperature, 6561/6579 pH, or 6565/6569
pH/ORP probes: Insert the probe into the
appropriate port, seating the pins of the two
Exo User Manual Chlorophyll Calibration Rhodamine Free
connectors. Push the probe inward so that it
seats into the port. Screw down the probe nut
by hand until it is finger tight. Use the long
end of the installation tool to tighten the probe
a quarter turn or until snug. Do not over-
tighten.
7. Ammonium, nitrate, and chloride ISE
probes do not have slip nuts and should be
installed without tools. Seat the probe body
directly on the sonde bulkhead. Use your
fingers to tighten.
!
!
!
!
!